Melting range determination of semi-crystalline polymers according to EN ISO 3146

BÜCHI Labortechnik AG

According to the official regulation EN ISO 3146, polymers do usually not have a sharp melting point as known for low-molecular substances. Here, we demonstrate the feasibility to employ the M-565 for the melting range determination of semi-crystalline polymers by capillary tube method according to EN ISO 3146.
Semi-crystalline poly(tetrafluoroethylene) (PTFE was chosen as test compound for the melting range measurements. For the analysed PTFE samples, we report a narrow melting range between 322.9 ± °C and 326.0 ± 0.1 °C.
With the M-565 the melting process can be measured and recorded without modification of the instrument

Stiahnuť aplikáciu Stiahnuť aplikáciu K prehľadu aplikácií Kontaktujte nás

Formulár na stiahnutie aplikácie

* Tieto polia sú povinné.

** Upozorňujeme, že vaše osobné údaje budú uložené bezpečne a v súlade so všeobecným nariadením o ochrane osobných údajov (GDPR) (EÚ) 2016/679 o ochrane osôb vo vzťahu k spracovaniu osobných údajov a voľnému pohybu týchto údajov. Máte právo požadovať od nás prístup k vašim uloženým osobným údajom, opraviť, vymazať alebo obmedziť spracovanie osobných údajov a kedykoľvek odvolať svoj súhlas. Osobné údaje nebudú zdieľané mimo našu skupinu spoločností ani predávané iným spoločnostiam. Osobné údaje môžu byť použité pre marketingové a komunikačné účely až do zrušenia súhlasu.