CZ  |  EN  |  SK

Melting range determination of semi-crystalline polymers according to EN ISO 3146

BÜCHI Labortechnik AG

According to the official regulation EN ISO 3146, polymers do usually not have a sharp melting point as known for low-molecular substances. Here, we demonstrate the feasibility to employ the M-565 for the melting range determination of semi-crystalline polymers by capillary tube method according to EN ISO 3146.
Semi-crystalline poly(tetrafluoroethylene) (PTFE was chosen as test compound for the melting range measurements. For the analysed PTFE samples, we report a narrow melting range between 322.9 ± °C and 326.0 ± 0.1 °C.
With the M-565 the melting process can be measured and recorded without modification of the instrument

Download application Download application Application overview Contact us

Application Download Form

* These fields are mandatory.

** Please note that your personal data will be stored safely and in accordance with the General Data Protection Regulation (GDPR) (EU) 2016/679 on the protection of individuals in relationship with processing personal data and the free movement of such data You have the right to request from us to access all your personal data stored, rectify it, delete or limit the processing of personal data, as well as to withdraw your consent at any time Personal data will not be shared outside our group of companies, nor sold to other companies Personal data may be used for marketing and communication purposes until consent is revoked

See more Applications in:

Analytics

See more Applications related to:

Measurement Instruments