CZ  |  EN  |  SK

Automated derivatization (HPTLC PRO)

Camag

Dimensions: 384 x 550 x 510 mm        Weight: ~ 35 kg 

General Data: 

  • Classification of installation and use: Class I, installation category II, protection class IPX0, Pollution degree 1 
  • Nitrogen or clean air pressure: 5 – 8 bar (73 – 116 psi) 
  • Operating temperature: 15 – 30 °C 
  • Recommended working temperature: 20 – 25 °C 
  • Humidity: Maximum relative humidity 80 % for temperatures up to 30 °C (Humidity and temperature conditions must not cause condensation!) 
  • Plate types: HPTLC Plates 20 x 10 cm, layer thickness 200 µm 

Electrical Data: 

  • Operating voltage: 100 - 240VAC; 50 / 60Hz 
  • Fluctuations ±10 % of the nominal voltage: Overvoltage category II 
  • Power consumption: 250 W 
  • Fuses: 4.0 AT, 250 VAC 
  • Max. supply current from power outlet: 8 A 

The HPTLC PRO Module DERIVATIZATION is part of the CAMAG® HPTLC PRO SYSTEM – the first fully automated HPTLC system worldwide. 

The HPTLC PRO Module DERIVATIZATION is designed for the fully automatic derivatization of HPTLC glass plates (20 x 10 cm) and uniquely combines two steps in a single device: high-precision spraying of derivatization reagents and heating of the plate. Employing the patented micro-droplet spraying technology, the Module DERIVATIZATION enables maximum homogeneity in applying derivatization reagents. The integrated plate heating unit ensures a uniform heat distribution across the plate. 

To suit the viscosity of the spraying reagents, four different nozzles are available. Spraying at slightly reduced pressure further improves the spraying process. Equipped with a fully automated nozzle changer and cleaning station, the Module DERIVATIZATION effectively avoids cross-contamination. 

Key Features: 

  • Part of the fully automated HPTLC PRO SYSTEM 
  • Nozzle changer to choose nozzles from three different nozzle parking positions 
  • Integrated plate heating unit 
  • Handling of two different reagents 
  • HPTLC glass plates (20 x 10 cm) 
  • Software-controlled by visionCATS 

Key Benefits: 

  • Spraying and heating in a single device 
  • Maximum homogeneity in reagent and heat distribution 
  • Highest user safety through automation and fume hood connection 
  • Optimal cleaning procedure between nozzle changes 
  • Low reagent consumption (3 mL) 

Contact us

Four different nozzles with different spraying properties for use depending on the derivatization reagents. The module is equipped with a nozzle changer and a fully automated nozzle cleaning station. Including 2 Carriers for HPTLC glass plates (20 x 10 cm) and a set of bottles for start-up.